text.skipToContent text.skipToNavigation

FXFF319 Flow Sensor with IO-Link

Usually ready to ship in 15 days after order is received.
The PG25 variant of the weFlux² flow sensors simultaneously measures the flow velocity and temperature of water, deionized water, and PG25. It is factory-calibrated for the water-glycol mixture used in liquid cooling, primarily in primary circuits at risk of freezing, but also in secondary circuits.

Sensor-specific data

Measuring Range 10 ... 400 cm/s
Temperature of the medium, flow measurement 0 ... 90 °C**
Temperature of the medium, temperature measurement 0 ... 90 °C
Setting Range  ≤ 400 cm/s
Medium PG25
Temperature measurement error ± 1 °C
MTTFd (EN ISO 13849-1) 1,210.41 a
Response time in case of temperature jump 10 s
Step response time T90 < 5 s

Environmental conditions

Ambient temperature -25 ... 80 °C
Storage temperature -25 ... 80 °C
Pressure Resistance 100 bar
EMC DIN EN 61326-1
Shock resistance per DIN IEC 68-2-27 30 g / 11 ms
Vibration resistance per DIN IEC 60068-2-6 5 g (10...2000 Hz)

Electrical Data

Supply Voltage 12 ... 32 V DC
Current Consumption (Ub = 24 V) < 40 mA
Number of Switching Outputs 2
Number of analog outputs 1
Analog Output 4...20 mA
Signal source Flow
Response Time 1 ... 5 s
Switching Output/Switching Current ± 100 mA
Switching Output Voltage Drop < 2 V
Load Current Voltage Output ≤ 20 mA
Operating delay time < 10 s
Short Circuit Protection yes
Reverse Polarity Protection yes
Protection Class III
Interface IO-Link V1.1

Mechanical Data

Setting Method IO-Link
Housing Material Stainless steel, V4A (1.4404/316L)
Material in contact with media Stainless steel, V4A (1.4404/316L)
Degree of Protection IP69K *
IP68 *
Connection Type Plug variant
Connection M12 × 1; 4-pin
Process Connection G 1/2"
Process Connection Length (PCL) 90.5 mm
Probe Length (PL) 50 mm
Rod diameter 6 mm
Weight 304 g

General Data

Scope of delivery 1 × initial start-up instructions
1 × sensor
Approvals FDA compliant

Output

Analog output flow yes
IO-Link yes
PNP NO yes

Adjustable parameters

Output Analog Output
Push-pull
NPN
PNP
Analog output 0...10 V
4...20 mA
Signal source Flow
Temperature
Circuit NC
NC/NO
NO
Other parameters 2 switching points
Selectable measurement range
All other parameters can be found in the product documentation.

Standards and certificates

B_SYM_CE_01.pdf B_SYM_CE_01.pdf B_SYM_CE_01.pdf B_SYM_CE_01.pdf B_SYM_CE_01.pdf B_SYM_CE_01.pdf
* Certified by wenglor
** The sensors were factory-calibrated for the PG25 medium and specified for this application. Technically, the sensors are suitable for a temperature of medium down to -25°C. To achieve a temperature below 0°C, the concentration must be increased. This results in a deviating measurement result, which is why use below 0°C must be tested individually for the specific mixture being used.

Connection Diagram

Dimensioned Picture

Dimensions specified in mm (1 mm = 0.03937")

Complementary Products

  • IO-Link converter
  • IO-Link Master
  • Software
Initial start-up instructions
Operating instructions
Interface protocol
Interface protocol
Interface_Protocol_IO-Link_Parameter_Process_Data_FXFF.pdf
Declarations of Conformity / Approvals
Product certificates
Product certificates
FXFF_IO-Link_V1.1_Manufacturer_declaration.pdf Fluid_sensors_food_safety_declaration.pdf